Hemolyzed, Clotted, Mislabeled: Preanalytical Errors in Blood Collection and How Competency Programs Reduce Them

A masked clinician at a computer workstation holds a filled blood collection tube beside a yellow rack of color-capped tubes

Photo: Gustavo Fring via Pexels

The phone call comes an hour after the draw. The sample is hemolyzed, the potassium cannot be reported, and the laboratory needs a recollection. The patient who was promised one stick gets a second one. The result the physician was waiting for arrives after the decision it was meant to inform. Nobody did anything dramatic. A tube went into a bag and came out unusable.

That is the most common form of laboratory error, and it does not happen in the laboratory. Three decades of measurement say most testing error occurs before the sample is analyzed, in the hands of the people who collect, label and transport it. For nurses, that makes specimen quality a competency rather than a laboratory problem.

This article covers where errors occur, why samples are rejected, what the standards and reviews identify as causes, what a recollection costs, how programs measure and reduce the rate, and what an assessor looks for. It is not a collection procedure; that is your laboratory's manual, taught against your facility's equipment.

Most laboratory error happens before the sample reaches the lab

Laboratory scientists divide testing into preanalytical, analytical and postanalytical phases. Mario Plebani, reviewing error surveys in Clinical Chemistry and Laboratory Medicine in 2006, found the preanalytical phase accounted for 46 to 68.2 percent of all errors, the postanalytical phase for 18.5 to 47 percent, and the analytical phase, the part inside the laboratory's instruments, for the smallest share.

His own laboratory's numbers show it. Carraro and Plebani (Clinical Chemistry, 2007) tracked a stat laboratory in Padua for three months: 160 confirmed errors in 51,746 analyses, 61.9 percent of them preanalytical against 15 percent analytical and 23.1 percent postanalytical. The two most common preanalytical causes were tube filling and patient identification. The overall rate had fallen since their 1996 audit, but the shape had not changed.

160 errors in 51,746 tests Preanalytical, 61.9%: before analysis, mostly at the bedside Postanalytical, 23.1%: reporting and interpretation Analytical, 15%: inside the laboratory Source: Carraro and Plebani, Clinical Chemistry 2007.
Published data from a single laboratory, not illustrative. Both leading preanalytical causes belong to the person holding the tube.

Sol Green's 2013 review in Clinical Biochemistry put the preanalytical share as high as 75 percent and estimated that about a quarter of those errors affect patient care. The part you control is where the risk sits.

The five rejection reasons, and what each one tells you

Laboratories reject samples for a short list of reasons, and the list is worth knowing because each item points to a different cause. Hemolysis, the rupture of red cells that releases their contents into the plasma, is the largest. Lippi and colleagues, writing in Clinical Chemistry and Laboratory Medicine in 2008, found hemolyzed specimens in up to 3.3 percent of routine samples and reported that hemolysis accounts for 40 to 70 percent of all unsuitable specimens, nearly five times the rate of the next causes combined.

Those next causes are clotting, where an anticoagulated sample has not stayed anticoagulated; underfilling, where the ratio of additive to blood is wrong because the tube did not draw to the volume its manufacturer specifies; labeling errors, where the tube is unlabeled, mislabeled or labeled away from the patient; and the wrong tube for the test ordered. A College of American Pathologists Q-Probes study across 147 laboratories (Wagar and colleagues, 2008) measured labeling errors at 0.92 per 1,000 labels, and an earlier Q-Probes study of 120 laboratories (Valenstein and colleagues, 2006) found that 85 percent of identification errors were caught before a result was released, which also means 15 percent were not.

Read the rejection, not just the request. A hemolyzed sample usually points at the draw or the transport. A clotted or underfilled one points at the tube handling. A labeling rejection points at the moment the label met the tube. Each one is diagnostic of a different habit.

Printable study guide

Get the printable Blood Draw Techniques study sheet

Blood Draw Techniques from Wahero Health Institute is a printable blood draw guide and techniques study sheet: a compact review to read beside your laboratory's collection manual and rejection criteria before a competency check.

Blood Draw Techniques, $10.99

Pay by Cash App, PayPal or Zelle at checkout; the download link is emailed once the payment is confirmed, usually the same day. Study material only: it does not replace the Board's requirements, your employer's policies or the judgment of a licensed clinician.

What the standards and the reviews say causes them

The collection standard in the United States is CLSI GP41, Collection of Diagnostic Venous Blood Specimens (seventh edition, 2017), and most facility manuals are written from it. It sets out the order in which tubes with different additives are collected, how tubes are filled and mixed, and how specimens are handled afterwards. The specific sequence, fill volumes and mixing instructions belong to the standard, the tube manufacturer and your laboratory, which is why they are not reproduced here. Identification has its own CLSI standard, PRE01, published in March 2024, and The Joint Commission's NPSG.01.01.01, carried into the hospital program's renamed National Performance Goals from January 2026 under a new number, requires specimen containers to be labeled in the presence of the patient.

On hemolysis specifically, the strongest evidence concerns the device. The CDC-funded Laboratory Medicine Best Practices review by Heyer and colleagues (Clinical Biochemistry, 2012) pooled 16 emergency department studies. All 11 that compared a new straight-needle venipuncture with a draw through an IV start found less hemolysis with the straight needle, a pooled risk ratio of 0.16, and the review rated the evidence high and recommended straight-needle venipuncture as a best practice. Four studies on site showed less hemolysis from the antecubital area, a risk ratio of 0.45. A separate meta-analysis by Lippi and colleagues (Biochemia Medica, 2013) reached the same conclusion about IV catheter draws.

Two practices with high-strength evidence for less hemolysis Pooled risk ratio for a hemolyzed sample; 1.0 means no difference, lower is less hemolysis Straight-needle venipuncture versus draw from an IV start, 11 studies Antecubital site versus other sites, 4 studies RR 0.16 RR 0.45 RR 1.0, no difference Source: Heyer et al., Clinical Biochemistry 2012;45(13-14):1012-1032. Laboratory Medicine Best Practices systematic review.
Published pooled estimates. The review recommends straight-needle venipuncture over IV-start draws as a best practice; whether and when a line draw is permitted at your facility is set by your policy and your laboratory.

What a recollection costs

The obvious cost is the patient's second needle. The less visible costs are the ones Green's review models: a repeat collection, a delayed result, a decision made on a stale or absent value, and in his estimate about 26 percent of preanalytical errors leading to unnecessary investigation, inappropriate treatment or a longer stay. A unit pays in nursing time twice, and the laboratory pays in instrument time and phone calls.

The identification error is the expensive one even when it is caught, because catching it means the laboratory stopped, investigated and recollected rather than reported. When it is not caught, a result belonging to one patient is acted on for another, which is the kind of event The Joint Commission's identification goal exists to prevent.

How programs measure and reduce preanalytical errors

Lippi's 2008 review set out the program most laboratories now run: continuous education for collectors, systematic detection and quantification of hemolysis in every sample using the analyzer's hemolysis index, a record of every non-conforming specimen, and a second specimen requested only for the tests hemolysis affects. The consequence for nurses is that the laboratory can report a hemolysis rate by unit, which turns a vague complaint into a number that can move.

Three interventions have evidence behind them. Barcoded labeling, reviewed by Snyder and colleagues in the same Laboratory Medicine Best Practices series (Clinical Biochemistry, 2012), was favored in all 17 included studies, with a pooled odds ratio of 4.39 for fewer identification errors and evidence rated high. Device choice, as above, carries the strongest hemolysis evidence. Training is the one with the honest caveat: Makhumula-Nkhoma and colleagues reviewed venipuncture training for hemolysis reduction in the Journal of Clinical Nursing in 2019 and found eight studies, none randomized; the four interventions that included mentoring reported falls in hemolysis of 0.4 to 19.8 percentage points, two reported small rises, and two were inconsistent. Training with feedback looks promising, but the controlled evidence does not yet exist.

The loop a specimen quality program runs Measure rejections by unit Feed back to the collectors Train, re-equip where the data points Measure again same definition Schematic. The measure comes from the laboratory's hemolysis index and rejection log; the definition must stay fixed between cycles.
Schematic, no data values. It follows the program Lippi and colleagues described in 2008: detect and quantify in every sample, record every non-conformity, educate continuously.

That loop lives in the competency validation program, and the rejection rate is what our piece on measuring whether nurse training worked calls a level-four outcome: a number the organization already collects and that changes when behavior changes.

A rejected sample is a data point, not a reprimand. Units that treat each rejection as an individual failing stop reporting them. Units that chart the rate by week and talk about the chart get the rate down.

What assessors look for

A specimen quality competency is assessed the way the errors occur, so an assessor watches the ends of the process more closely than the middle: identification against two identifiers and labeling in the patient's presence, because those are The Joint Commission's gates; tube choice and order taken from the laboratory's chart rather than memory; and whether each tube fills to its designed volume and is mixed the way the chart says, because underfilling and clotting are the next two rejections on the list.

They watch device choice against policy: whether you draw through a line when the policy says venipuncture, and whether you can say why the policy says it. They watch handling after the draw, because time, temperature and agitation in transport are preanalytical variables your laboratory sets limits on. And they watch what you do with a doubtful sample. Saying that a tube is short and needs recollecting is a pass. Sending it is a fail, and the laboratory's record will show it was.

Our printable Blood Draw Techniques study sheet is written to be read beside your laboratory's rejection criteria for exactly this reason. The same instinct runs through what IV insertion competency really takes: the decision is the skill, and the puncture is only the part that frightens people.

Educational use. This article is learning material for nurses and nursing students. It is not clinical advice, and it does not replace your employer's policies, your facility's protocols, or the judgement of a licensed clinician. Always follow the standards and procedures in force where you practise.

Key takeaways

Frequently asked questions

What counts as a preanalytical error? Anything that goes wrong between the test being ordered and the sample being analyzed: wrong patient, wrong tube, wrong label, underfilled or clotted or hemolyzed sample, and delays or poor handling in transport. Plebani's 2006 review in Clinical Chemistry and Laboratory Medicine put this phase at 46 to 68 percent of all laboratory errors, and most of it happens at the bedside rather than in the laboratory.

Why do samples drawn through an IV catheter hemolyze more often? The CDC-funded Laboratory Medicine Best Practices review by Heyer and colleagues (2012) found that every one of the 11 studies comparing the two methods showed lower hemolysis with a new straight-needle venipuncture than with a draw from an IV start, a pooled risk ratio of 0.16, and a 2013 meta-analysis by Lippi and colleagues reached the same conclusion. Whether a line draw is permitted, and when, is set by your facility's policy and your laboratory; the competency is knowing that policy and the evidence behind it.

Does training reduce hemolysis rates? The evidence is mixed. A 2019 systematic review in the Journal of Clinical Nursing by Makhumula-Nkhoma and colleagues found eight studies, none randomized; four interventions that included mentoring reported falls in hemolysis of 0.4 to 19.8 percentage points, two reported small rises, and two were inconsistent. Training with mentoring looks promising, but the controlled evidence does not yet exist, which is one reason programs measure their own rates.

Measure the rate, then train to it

A specimen quality program that starts with a lecture and no number has nothing to show at the end. Wahero Health Institute builds collection competency validation on site, against your laboratory's manual and rejection criteria, with the laboratory's own rejection rate as the measure of whether it worked.

IV Therapy and Vascular Access Course Schedule Facility Training

Educational content only

This material is published by Wahero Health Institute for professional education and is not individual medical advice, a care protocol, or a substitute for clinical judgment. Always follow your facility's policies, your state's nurse practice act, and your own scope of practice, and confirm medication doses against a current authoritative reference before administration. See our Terms of Use.